The Journal of the American Dental Association
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Am Dent Assoc, Vol 137, No 9, 1275-1280.
© 2006 American Dental Association

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Chin, J. R.
Right arrow Articles by Palenik, C. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chin, J. R.
Right arrow Articles by Palenik, C. J.
Related Collections
Right arrow Infection Control

RESEARCH

JADA Continuing Education

Internal contamination of air-driven low-speed handpieces and attached prophy angles



Judith R. Chin, DDS, MS, Chris H. Miller, MS, PhD and Charles John Palenik, MS, PhD, MBA

Background. In an in vitro crossover study, the authors investigated whether the interior of low-speed handpiece/prophy-angle systems becomes contaminated during operation and submersion into Geobacillus stearothermophilus.

Methods. This study involved two types of handpieces attached to eight brands of prophy angles. The researchers operated angles attached to sterile handpieces for 60 seconds. They then analyzed the inside surfaces of the angle, nosecone and motor. They tested each prophy angle and handpiece 10 times.

Results. In the 160 tests of handpieces contaminated at the prophy cup end, the spores traveled into the motor gears 32 times (20 percent). In the other 160 tests in which the motor gears were contaminated, the test bacterium traveled through the prophy cup in 75 instances (47 percent).

Conclusions. The in vitro data suggest that low-speed handpiece motors can become contaminated internally during use with prophy angles. Also, internal contaminants appear to have been released from the handpiece.

Clinical Implications. The results suggest that low-speed hand-pieces can become contaminated internally during use. Unless low-speed handpieces are sterilized properly between patients, they may become cross-contaminated.

Key Words: Infection control; handpiece; prophy angle; Geobacillus stearothermophilus; equipment contamination; sterilization