The Journal of the American Dental Association
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


J Am Dent Assoc, Vol 130, No 9, 1301-1311.
© 1999 American Dental Association

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by LEONARD, D. L.
Right arrow Articles by CHARLTON, D. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by LEONARD, D. L.
Right arrow Articles by CHARLTON, D. G.
Related Collections
Right arrow Dental Equipment/Instruments

RESEARCH

JADA Continuing Education

PERFORMANCE OF HIGH-SPEED DENTAL HANDPIECES SUBJECTED TO SIMULATED CLINICAL USE AND STERILIZATION



DANIEL L. LEONARD, D.D.S. and DAVID G. CHARLTON, D.D.S., M.S.D.

Background. The authors investigated the performance of nine commercially available high-speed air-turbine dental handpieces subjected to 1,000 simulated clinical uses and sterilizations.

Methods. Six new handpieces from each of nine different models were subjected to simulated clinical use with a custom-made handpiece wear tester and then autoclaved. Ten parameters related to clinical performance (longevity, power, turbine speed, fiberoptic transmission, eccentricity, noise, chuck performance, visibility angle, interocclusal clearance and water coolant spray pattern) were measured at baseline and after 250, 500, 750 and 1,000 use/sterilization cycles.

Results. Power, turbine speed, eccentricity and noise performance were statistically analyzed using one-way analysis of variance and Tukey post hoc pairwise comparison tests at the .05 significance level. At baseline, significant differences were found between models for all of these parameters. In general, from baseline to 1,000 cycles, the handpieces exhibited greater eccentricity and reduced fiberoptic performance. Longevity data analyzed by using Gehan’s generalized Wilcoxon test for comparison of survival distributions ({alpha} = .05) revealed significant differences between the handpiece models.

Conclusions. The results of this study indicate that no handpiece model is superior to the others in all parameters evaluated. All models evaluated can be expected to perform for at least 500 clinical use/sterilizations, or approximately one year, if properly maintained.

Clinical Implications. Clinicians need to be able to identify handpieces that can withstand repeated heat sterilization without loss of performance or longevity. The results of this study will aid clinicians in selecting handpiece models that meet their needs.